You are viewing the site in preview mode
Skip to main content
|
Effective compression depth (mm)
|
Mean (95% confidence interval)
|
|---|
|
Source of feedback
| |
|---|
| |
Anterior sensor
|
Internal measure
|
Difference
|
|---|
|
Typical mattress only
|
37.8 (34.1, 41.5)
|
52.1 (50.1, 54.1)
|
14.3 (12.0, 16.5)
|
|
Typical mattress + backboard
|
42.9 (39.5, 46.3)
|
51.7 (49.8, 53.5)
|
8.7 (6.5, 11.0)
|
|
Memory foam mattress only
|
41.2 (37.9, 44.4)
|
54.2 (52.3, 56.1)
|
13.0 (10.7, 15.3)
|
|
Memory foam mattress + backboard
|
46.3 (43.2, 49.3)
|
53.7 (52.0, 55.5)
|
7.5 (5.2, 9.8)
|
- Presented are the mean adjusted values from a mixed effects linear model. Fixed effect: mattress type, CPR backboard use, feedback source, interaction between feedback source and mattress type, and interaction between feedback source and CPR backboard use. Random effect: intercept, mattress type, CPR backboard use, and feedback source. Goodness of fit: marginal R
2 = 0.43, conventional R
2 = 0.80